Salt spray corrosion test chamber
Product function and purpose
The equipment uses thermal radiation heating to maintain the required temperature of the space to ensure test accuracy.
Product specifications
Test machine type number |
EHG-YW1000 |
EHG-YW2000 |
EHG-YW4000 |
Inner dimensions (cm) | 100 * 100 * 100 | 120 * 120 * 139 | 150 * 145 * 184 |
Content product (L) |
1000 |
2000 |
4000 |
Lab temperature range (c) | RT+5 to 60 | ||
Saturated barrel temperature range (c |
RT ~ 70 |
||
Humidity range (RH |
P 90 | ||
Saltwater spray amount | 1 ~ 2ml/80cm/h(average for at least 16 hours of spray) | ||
Spray pressure (KPa) |
70 to 170 |
||
Potion PH | 6.5 to 7.2/3.0 to 3.2 | ||
Collect liqu id PH |
6.5 to 7.2/3.1 to 3.3 | ||
Sample placement Angle | 20±5。(angled to the vertical plane) | ||
Control accuracy (C temperature deviation: ±2.0 |
Testing machine standard
1. It meets the requirements of GB / t2611-2007 general technical requirements for testing machines, GB / t16826-2008 electrohydraulic servo universal testing machines and JB / t9379-2002 technical conditions for tension compression fatigue testing machines;
2. Meet GB / t3075-2008 metal axial fatigue test method, GB / t228-2010 metallic materials tensile test method at room temperature, etc;
3. It is applicable to GB, JIS, ASTM, DIN and other standards.